A serial dilution spreads a single stock across many tubes, each diluted by the same factor from the one before it. This calculator shows the concentration at every tube, the cumulative dilution factor, and the exact transfer and diluent volumes so your series is reproducible.
How it works
Each step divides the previous concentration by the same dilution factor, so the concentrations fall geometrically:
concentration[n] = stock / DF^n (n = 0 is the undiluted stock)
cumulative DF[n] = DF^n
transfer volume = total volume / DF
diluent volume = total volume − transfer volume
A 1:10 (DF = 10) series therefore produces concentrations of stock, stock/10, stock/100, stock/1000, and so on. For 1 mL tubes you move 0.1 mL forward into 0.9 mL of fresh diluent at every step.
Worked example
Starting from a 1,000 µg/mL stock with a dilution factor of 10 over five steps in 1 mL tubes:
| Tube | Concentration | Cumulative DF | Transfer | Diluent |
|---|---|---|---|---|
| Stock (0) | 1,000 µg/mL | 1× | — | — |
| 1 | 100 µg/mL | 10× | 0.1 mL | 0.9 mL |
| 2 | 10 µg/mL | 100× | 0.1 mL | 0.9 mL |
| 3 | 1 µg/mL | 1,000× | 0.1 mL | 0.9 mL |
| 4 | 0.1 µg/mL | 10,000× | 0.1 mL | 0.9 mL |
| 5 | 0.01 µg/mL | 100,000× | 0.1 mL | 0.9 mL |
Counting CFU from plated dilutions
For bacterial plate counts, the goal is to find the tube whose plated dilution lands in the 30–300 colony countable range. From that plate:
CFU/mL of original stock = colony count / (volume plated × cumulative DF)
For example, if 0.1 mL from Tube 4 (cumulative DF = 10,000) gives 85 colonies:
CFU/mL = 85 / (0.1 mL × 10,000) = 85,000 CFU/mL
Only count plates in the 30–300 range; fewer than 30 colonies gives statistically unreliable counts, and more than 300 become too crowded to distinguish.
Common sources of error
Reused pipette tips: residual liquid on a tip carries concentrated sample into the next tube, inflating that dilution. This is the most common cause of curved calibration lines that should be linear, and of over-counted plates. Always use a fresh tip and vortex or pipette-mix each tube before the next transfer.
Incomplete mixing: transfer 0.1 mL into 0.9 mL and mix thoroughly before drawing the next transfer. Pulling from an unmixed tube takes sample from whatever concentration zone the pipette tip enters.
Wrong diluent volume: if your total tube volume is not what you entered in the calculator, the transfer fraction changes. For a 2 mL tube with DF = 10, transfer is 0.2 mL into 1.8 mL, not 0.1 mL into 0.9 mL.
Losing track of the stock: always label Tube 0 as the undiluted stock so the numbering stays unambiguous when you come back to the rack 30 minutes later.
Tips
- Mix each tube thoroughly before every transfer.
- Use a fresh pipette tip per step — carryover is the most common source of error.
- For dilutions spanning more than 6 orders of magnitude, consider preparing two overlapping series from the stock and the intermediate tube to reduce cumulative pipetting error.